CN-0288 Software User Guide
Overview
CN-0288 is a linear variable differential transformer (LVDT) signal conditioning circuit. The system converts the output of the LVDT sensor to a unipolar DC voltage and drives the inputs of a successive approximation analog-to-digital converter (ADC).
This circuit uses a single chip solution that drives the excitation signal of the primary side of the LVDT and converts the secondary side. The complete system uses a 12-bit ADC resulting in a dynamic range of 82dB at a standard LVDT bandwidth of 250Hz. (The term bandwidth used in this case is referring to a sinusoidally varying mechanical position and can be easily configured to meet wider bandwidth applications by configuring a few external passive components.)
The signal conditioning circuitry of the system consumes only 15mA of current from the ±15V supply and 3mA from the +5V supply making this configuration suitable for industrial precision position and gauging applications such as flight control surface position feedback.
This circuit note discusses the design steps needed to optimize the circuit shown in Figure 1 for a specific bandwidth including noise analysis and component selection considerations.
This user guide will discuss how to use the evaluation software to collect data from the EVAL-CN0288-SDPZ Evaluation Board (CN-0288 Board)
Required Equipment
- EVAL-SDP-CB1Z Controller Board (SDP-B Board)
- EVAL-CN0288-SDPZ Evaluation Board (CN-0288 Board)
- EVAL-CFTL-6V-PWRZ (+6V Power Supply) or equivalent
- EVAL-CFTL-LVDT Linear Variable Differential Transformer (Measurement Specialties, Inc. E-100 Economy Series LVDT)
-
- (supplied with provided CD in kit)
- PC with the following Minimum Requirements
- USB type A to USB type mini-B cable
General Setup
- The EVAL-CN0288-SDPZ (CN-0288 Board) connects to the EVAL-SDP-CB1Z (SDP-B Board) via the 120-Pin connector
- The EVAL-CFTL-6V-PWRZ (+6V DC Power Supply) powers the EVAL-CN0288-SDPZ (CN-0288 Board) via the DC barrel jack
- The EVAL-SDP-CB1Z (SDP-B Board) connects to the PC via the USB cable.
- The EVAL-CFTL-LVDT (LVDT) connects to the EVAL-CN0288-SDPZ (CN-0288 Board) via the 6-pin header at J3.
Installing the Software
- Extract the file CN0288 Eval Software.zip and open the file setup.exe.
NOTE: It is recommended that you install the CN-0288 Evaluation Software to the default directory path C:/Program Files/Analog Devices/CN0288/ and all National Instruments products to C:/Program Files/National Instruments/
- Click Next to view the installation review page
- Click Next to start the installation
- Upon completion of the installation of the CN-0288 Evaluation Software, the installer for the ADI SDP Drivers will execute.
NOTE: It is recommended that you close all other applications before clicking “Next”. This will make it possible to update relevant system files without having to reboot your computer.
- Press “Next” to set the installation location for the SDP Drivers.
It is recommended that you install the drivers to the default directory path
C:/Program Files/Analog Devices/SDP/Drivers
- Press “Next” to install the SDP Drivers and complete the installation of all software. Click “Finish” when done.
Connecting the Hardware
- Connect the EVAL-CFTL-LVDT (LVDT) to J3 of the EVAL-CN0288-SDPZ (CN-0288 Board) as depicted below.
NOTE: If a different LVDT is used other than the E-100 LVDT from Measurement Specialties, Inc the wiring schematic will be different!
- Connect the EVAL-CFTL-6V-PWRZ (+6V DC Power Supply) to the Barrel Jack at J4 or the Screw Terminal at J1 on the EVAL-CN0288-SDPZ (CN-0288 Board)
Using the Evaluation Software
Software Control and Indicator Descriptions
- Connect/Reconnect Button
- When this button is pressed, the SDP-B Board makes a USB connection to the CN-0288 Board. A connection to the SDP-B Board must be made to use the software.
- Run Button
- When this button is pressed, the SDP-B Board will collect conversion data and present the acquisitions in the chart.
- Stop Button
- When this button is pressed, the software stop collecting data from the CN-0288 Board
- Save Data Button
- When this button is pressed, the software will save the data collected to a tab delimited ASCII spreadsheet file.
- Clear Data Button
- When this button is pressed, the software will clear all data collected from the chart history.
- Control Tabs
- Acquire Data - Clicking this tab brings the data collection chart to the front.
- Configure System - Clicking this tab bring the system configuration settings to the front.
- Channel 1 Numerical Indicator
- This indicator displays the current position from null on Channel 1.
- Channel 2 Numerical Indicator
- This indicator displays the current position from null on Channel 2.
- Channel Enable Checkboxes
- This control determines which channels display in the chart.
- Select Units Radio Buttons
- This control determines the Y-Axis units of the data displayed in the chart.
- ADC Conversion Out of Range LED Indicator
- This indicator determines whether a conversion result is outside the allowable range of conversion results.
- Chart Controls
- These controls allow the user to zoom-in, zoom-out, and pan through the data collected.
- System Status String Indicator
- This indicator displays a message to the user detailing the current state of the software.
- System Status LED Indicator
- This indicator displays the current state of the software in the form of an LED. There are three status LED colors.
Inactive
Busy
Error
- ADC Conversion Thresholds Panel
- This panel allows the registers in the AD7992 analog-to-digital converter to be written to and read back.
- LVDT Full Scale Range Panel
- This panel configures the maximum stroke range of the LVDT. The default configuration is 0.2 inches.
Establishing a USB Connection Link
- Follow the instructions to properly install the software and connect the hardware as described in the previous sections.
- Open the file named CN0288.exe in the installation directory.
NOTE: If the software was installed to the default location it will be found at
C:/Program Files/Analog Devices/CN0288/CN0288.exe
- Click the Connect/Reconnect Button. A window with a progress bar will load.
- Upon success, the System Status String Indicator will display SDP Board Ready to Acquire Data
Capturing a Data
- Establish a USB Connection Link.
- Click the Run Button
- Click the Stop Button when acquisition is complete.
Saving Data to a Spreadsheet File
- Establish a USB Connection Link.
- Capture Data.
- Click the Save Data Button.
- Configure the format of the spreadsheet data.
- Select either Inches or ADC Codes.
- Select either Channel 1, Channel 2, or select both channels.
- Click the OK Button.
- Browse to the directory location where the spreadsheet file is to be saved.
- Name the file.
- Click the OK Button.
The software saves the spreadsheet file as ASCII text with columns separated by tabs.
Setting ADC Conversion Thresholds
- Establish a USB Connection Link.
- Click the “Configure System” Control Tab.
- Change the hexadecimal value of the register to be written in the ADC Conversion Thresholds Panel.
- Press the Set Thresholds Button.
Information regarding the internal register structure can be found in the AD7992 Datasheet.
Reading ADC Conversion Thresholds
- Establish a USB Connection Link.
- Click the “Configure System” Control Tab.
- Press the Read Thresholds Button.
The threshold contents will be displayed in the ADC Conversion Thresholds Panel.
Setting the Full Scale LVDT Range
- Click the “Configure System” Control Tab.
- Change the value of the channel of interest in the LVDT Full Scale Range Panel.
- The default value is 0.20 inches for both channels.
- CN0335软件用户指南
- CN0336软件用户指南
- CN0349软件用户指南
- CN0350软件用户指南
- CN-0354软件用户指南
- CN0355软件用户指南
- CN0196软件用户指南
- CN0203软件用户指南
- CN0204软件用户指南
- CN-0269软件用户指南
- CN-0271软件用户指南
- CN-0296软件用户指南
- CN-0301软件用户指南
- CN0288 设计和集成文件
- 基于CN-0288LVDT Conditioning的参考设计
- 上位机软件的用户界面设计技巧 773次阅读
- 搞不懂组态软件和HMI?常见组态软件有哪些? 4789次阅读
- 常用的远程控制软件介绍 1763次阅读
- 医疗器械软件风险分析介绍 医疗设备软件静态和动态分析技巧 1850次阅读
- 如何使用openEuler用户软件仓(EUR) 1223次阅读
- 基于STM32 MCU应用的EMC指南之软件篇 1020次阅读
- 开源、可重用的软件堆栈支持CbM的实时处理和算法开发 900次阅读
- EsDA为用户带来简单易用的软件设计体验 1884次阅读
- 使用含快速开关SiC器件的RC缓冲电路实用解决方案和指南 2531次阅读
- 常见的组态软件有哪些 1.2w次阅读
- 如何正确的设置IAR单片机编程软件 4994次阅读
- 一款以BitTorrent代币连接和奖励用户的软件BitTorrent Speed介绍 3328次阅读
- 组态软件的特点和应用 1.4w次阅读
- 组态软件是什么_组态软件有什么用 3.5w次阅读
- 长虹CN-7机芯彩电场扫描电路分析 1w次阅读
下载排行
本周
- 1电子电路原理第七版PDF电子教材免费下载
- 0.00 MB | 1491次下载 | 免费
- 2单片机典型实例介绍
- 18.19 MB | 95次下载 | 1 积分
- 3S7-200PLC编程实例详细资料
- 1.17 MB | 27次下载 | 1 积分
- 4笔记本电脑主板的元件识别和讲解说明
- 4.28 MB | 18次下载 | 4 积分
- 5开关电源原理及各功能电路详解
- 0.38 MB | 11次下载 | 免费
- 6100W短波放大电路图
- 0.05 MB | 4次下载 | 3 积分
- 7基于单片机和 SG3525的程控开关电源设计
- 0.23 MB | 4次下载 | 免费
- 8基于AT89C2051/4051单片机编程器的实验
- 0.11 MB | 4次下载 | 免费
本月
- 1OrCAD10.5下载OrCAD10.5中文版软件
- 0.00 MB | 234313次下载 | 免费
- 2PADS 9.0 2009最新版 -下载
- 0.00 MB | 66304次下载 | 免费
- 3protel99下载protel99软件下载(中文版)
- 0.00 MB | 51209次下载 | 免费
- 4LabView 8.0 专业版下载 (3CD完整版)
- 0.00 MB | 51043次下载 | 免费
- 5555集成电路应用800例(新编版)
- 0.00 MB | 33562次下载 | 免费
- 6接口电路图大全
- 未知 | 30320次下载 | 免费
- 7Multisim 10下载Multisim 10 中文版
- 0.00 MB | 28588次下载 | 免费
- 8开关电源设计实例指南
- 未知 | 21539次下载 | 免费
总榜
- 1matlab软件下载入口
- 未知 | 935053次下载 | 免费
- 2protel99se软件下载(可英文版转中文版)
- 78.1 MB | 537793次下载 | 免费
- 3MATLAB 7.1 下载 (含软件介绍)
- 未知 | 420026次下载 | 免费
- 4OrCAD10.5下载OrCAD10.5中文版软件
- 0.00 MB | 234313次下载 | 免费
- 5Altium DXP2002下载入口
- 未知 | 233046次下载 | 免费
- 6电路仿真软件multisim 10.0免费下载
- 340992 | 191183次下载 | 免费
- 7十天学会AVR单片机与C语言视频教程 下载
- 158M | 183277次下载 | 免费
- 8proe5.0野火版下载(中文版免费下载)
- 未知 | 138039次下载 | 免费
评论
查看更多