In all areas of our daily life, we rely upon clean products that are not contaminated with pathogenic germs. This is particularly important for medical diagnostics and medical products. Equipment like surgical instruments, injection machines or endoscopes has to be disinfected before usage.
Although in-vitro diagnostics are not applied directly to patients but cover analysis of blood, urine or other bodily
substances, germ-free equipment is necessary. If micro-organisms contaminate laboratory equipment and test tubes, diagnostic results would be adulterated or wasted.
Regulations, standards and laws define sterilisation requirements. The respective public authorities, like the Food and Drug Administration (FDA) in the USA, control compliance with these requirements.
How to sterilise
There are different methods of achieving a sterile product or instrument. The most common way to kill micro-organisms like bacteria, fungi and viruses is by exposure to heat. For medical purposes, autoclaves are widely used. Within an autoclave, objects to be sterilised are exposed to steam at a temperature of around 125°C for a set period of time. The high temperature, humidity and pressure of the steam are real disadvantages of this approach. Some materials cannot withstand these conditions.
For this reason, another method is increasingly applied for sterilisation: Irradiation by gamma or X-rays. This approach offers several advantages. Since gamma rays are highly penetrating, the goods can be disinfected after packaging, within bigger boxes or even on pallets at room temperature. That is to say, the objects are sterilised at the last stage of the production process, and costly high-class clean-room conditions during production might not be necessary.
The irradiation dose depends on the original microbial condition of the goods and the target safety level. Usually it is required to apply a dose of 25kGy to 45kGy – a kiloGray being the Standard International unit of absorbed radiation dose of ionising radiation.
How to track sterilised products
Modern logistic processes use Radio Frequency Identification (RFID) technology to label and track objects. In the simplest form, an electronic RFID tag contains a unique ID, similar to a barcode label. In contrast to the latter, RFID tags can be read contactless, via an air interface without direct optical access to the tag. If several products are packed into a bigger box, it is possible to identify each single object without opening or unpacking the outer box.
Even more important, the electronic RFID labels allow the storage of additional data onto the tags. Data can be written during the manufacturing process (e.g. lot number, manufacturing date, expiry date, product type, etc.) and during the flow through the supply chain (e.g. leaving the factory, date of sterilisation, name of logistic partner, etc.)
For medical instruments and products in particular, it is very important to reliably identify the objects as well as document the treatment and manufacturing history. Mistakes can lead to serious problems and risk for the life of patients.
To leverage these features to the full extent for medical applications, it is necessary to combine the gamma ray irradiation and RFID technologies already mentioned.
Conventional E2PROM based RFID products do not withstand the irradiation process. After exposure to the irradiation, the memory content is erased. At this point, Fujitsu’s FRAM technology comes into play.
FRAM technology
In contrast to the conventional non-volatile memories, Flash and E2PROM, the content of an FRAM cell is not stored in the form of charge carriers in a ‘floating gate’。 The information – logically 0 or 1 – is contained in the polarisation of the ferroelectric material lead zirconate titanate, PZT (Pb (ZrTi)O3)。 This material is placed between two electrodes in the form of a thin film, in a similar way to the structure of a capacitor. When an electrical field is applied the material polarises in one direction and retains this structure even after the field has been removed. If the direction of the electrical field is reversed, the atoms polarise accordingly in the opposite direction. An FRAM memory cell has the same structure as a DRAM cell and consists of a transistor and a capacitor, but in this case the FRAM cell contains a capacitor with a ferroelectric dielectric.
The energy, applied during an irradiation process, removes the charge in floating gates of E2PROM cells but does not affect the polarisation of FRAM cells. Scientific studies have proven FRAM’s resistance to irradiation, for doses up to 50kGy.
Further FRAM advantages FRAM offers further advantages,
especially for RFID products. Since no large charge quantities have to be displaced, charge pumps to generate high programming voltages are not necessary. Consequently, FRAM technology is much more energy-saving than E2PROM. This directly affects the operating range of
20 epd
september 2011 medical electronics
FRAM information – logically 0 or 1 – is contained in the polarisation of the ferroelectric material lead zirconate titanate, PZT (Pb (ZrTi)O3)
based RFID chips, Fujitsu’s FerVID family enables the same high speed data transfer rate for both reading and writing over long distances. The write endurance, specified to 10 billion cycles, is far higher than that of conventional RFID tags, thus saving cost and time in applications. Fujitsu offers HF and UHF FRAM RFID products. The ISO 15693 compliant devices cover the 13.56MHz operating frequency and feature 2kByte and 256Byte memories respectively. In the UHF band of 860 to 900MHz, Fujitsu offers the 4kByte FRAM device, compliant to the EPC global C1G2 standard. The products feature anti-collision, lump and fast read/write transmission commands.
RFID tags in a positive way. Due to the low power consumption nature of FRAM, the operating range is higher for a given field strength or power density.
FRAM memories can be written as fast as they can be read, i.e. FRAM’s write access is about 25 times faster than E2PROM’s write access. The maximum number of write/delete cycles for Flash and E2PROM is between 10,000 and 100,000. If this limit is exceeded, the memory content can no longer be reliably
stored. By comparison with over 10 thousand million write/delete cycles (1010 ), the lifetime of an FRAM memory is almost unlimited. This allows the re-use of FRAM tags many times over. Occasionally, FRAM is incorrectly associated with ferromagnetism. Magnetic fields do not affect the ferroelectric material.
Summary
Compared to conventional E2PROM/flash
Furthermore, a dual-interface device is also available in two different types – one as a conventional contactless EPCglobal RFID product and the other as a derivative with an additional contact-based SPI interface. This dual-interface type can be implemented as part of a Microcontroller-based embedded system.
DIRK FISCHER, Product Marketing
Engineer, Business Unit Automotive &
Embedded. Fujitsu
QHighCopper
capacitorscapacitorsBarrier
capacitorsTCC/VCCMLCCsSafety-Non250Vac
+L 5HOSURGXFWVÊ SpaceMLCCsFlexiCap™
Ê IECQ-CECC
Ê AEC-Q200
6DIHW\&HUWL¿HG0/&&V
- 《医用电子直线加速器》pdf 0次下载
- 医用电器环境要求及试验方法综述 13次下载
- 从FRAM迁移到EERAM 7次下载
- 软件无线电RFID测试平台课件下载 32次下载
- RFID防碰撞技术的详细资料研究说明 65次下载
- 超高频无源RFID标签电路分析如何提高恢复电路效率
- 如何在RFID中嵌入FRAM
- RFID读写器天线的工作原理和设计的基本步骤说明 83次下载
- RFID技术中不可避免的碰撞问题应该如何解决 31次下载
- RFID技术的基本工作原理和设计RFID读写器天线的资料说明 45次下载
- 中国射频识别RFID技术政策的详细资料说明 17次下载
- RFID成为自动化控制发展新趋势 7次下载
- 基于单片机的医用灭菌器控制系统设计 19次下载
- 医用超声诊断系统中双波束合成失真的校正 35次下载
- 基于单片机的某医用灭菌器控制系统开发
- MRAM与FRAM技术有何差异 1591次阅读
- 地下管网的管理如何利用rfid技术 584次阅读
- 电网资产管理加入RFID技术会有什么效果 739次阅读
- 微流控技术如何有效提高肿瘤化疗效果 2393次阅读
- 射频RFID信号攻击模拟效果探讨 1549次阅读
- 无菌医疗器械生产有哪些常用的灭菌方法?如何进行验证? 9690次阅读
- 对于医用灭菌器中的压力测量,如何选择正确适当的压力传感器 1722次阅读
- 详解医用灭菌器控制系统 2746次阅读
- RFID技术提高仓库管理工作效率简析 1356次阅读
- RFID技术原理、特点及应用详解 2w次阅读
- 一种基于ZigBee技术的RFID系统网络构建方法 2889次阅读
- 基于ZigBee技术的有源RFID系统 2059次阅读
- 基于RFID与车牌识别技术相结合的管理方法 1247次阅读
- 浅析RFID技术在医院机构的使用案例 1434次阅读
- 在嵌入式设计中将FRAM用作闪存的替代方案 1933次阅读
下载排行
本周
- 1电子电路原理第七版PDF电子教材免费下载
- 0.00 MB | 1491次下载 | 免费
- 2单片机典型实例介绍
- 18.19 MB | 95次下载 | 1 积分
- 3S7-200PLC编程实例详细资料
- 1.17 MB | 27次下载 | 1 积分
- 4笔记本电脑主板的元件识别和讲解说明
- 4.28 MB | 18次下载 | 4 积分
- 5开关电源原理及各功能电路详解
- 0.38 MB | 11次下载 | 免费
- 6100W短波放大电路图
- 0.05 MB | 4次下载 | 3 积分
- 7基于单片机和 SG3525的程控开关电源设计
- 0.23 MB | 4次下载 | 免费
- 8基于AT89C2051/4051单片机编程器的实验
- 0.11 MB | 4次下载 | 免费
本月
- 1OrCAD10.5下载OrCAD10.5中文版软件
- 0.00 MB | 234313次下载 | 免费
- 2PADS 9.0 2009最新版 -下载
- 0.00 MB | 66304次下载 | 免费
- 3protel99下载protel99软件下载(中文版)
- 0.00 MB | 51209次下载 | 免费
- 4LabView 8.0 专业版下载 (3CD完整版)
- 0.00 MB | 51043次下载 | 免费
- 5555集成电路应用800例(新编版)
- 0.00 MB | 33562次下载 | 免费
- 6接口电路图大全
- 未知 | 30320次下载 | 免费
- 7Multisim 10下载Multisim 10 中文版
- 0.00 MB | 28588次下载 | 免费
- 8开关电源设计实例指南
- 未知 | 21539次下载 | 免费
总榜
- 1matlab软件下载入口
- 未知 | 935053次下载 | 免费
- 2protel99se软件下载(可英文版转中文版)
- 78.1 MB | 537793次下载 | 免费
- 3MATLAB 7.1 下载 (含软件介绍)
- 未知 | 420026次下载 | 免费
- 4OrCAD10.5下载OrCAD10.5中文版软件
- 0.00 MB | 234313次下载 | 免费
- 5Altium DXP2002下载入口
- 未知 | 233046次下载 | 免费
- 6电路仿真软件multisim 10.0免费下载
- 340992 | 191183次下载 | 免费
- 7十天学会AVR单片机与C语言视频教程 下载
- 158M | 183277次下载 | 免费
- 8proe5.0野火版下载(中文版免费下载)
- 未知 | 138039次下载 | 免费
评论
查看更多